Stability test chamber 700 lt - FSC 700
Stability Chambers without humidity control - Horizontal Airflow
Stability Chamber FSC 700 is 700 lt instrument for high technology laboratory testing the reaction of samples at several environmental conditions at more temperatures. Before launching a product on the market, it is necessary to verify how it reacts at high or low temperatures.
The model FSC has not humidity control and it has horizontal forced ventilation to test many small samples on more shelves; this chamber is suggested for food, pharma, cosmetics industries or some fields where samples have small dimensions.
700 lt capacity makes this instrument like a big, refrigerated incubator with better results, granted by good uniformity and accuracy.
FIELDS OF APPLICATION
- Pharmaceutical industry with ICH test
- Laboratory testing
- Quality test
- Chemical industry
- Environmental test
- Food industry
- Cosmetic industry
STANDARD FEATURES
- Cooling system controller by solenoid electrovalves
- Safety thermostat, to protect the samples against high temperatures (optionally low temperatures)
- System setting protected by password
- Key lock on the door
- Temperature range from +4°C to +60°C (optionally from -10/-20/-30°C to +60°C without humidity control)
- Models include forced air refrigeration (distributed uniformly vertically), and have high temperature homogeneity inside of the chamber
- Ultrasound humidity generator, introducing microscopic water drops inside of the chamber at room temperature (electrode system optionally)
- Condensation drying system by cooling
- Independent cooling and heating systems
- Tropicalization treatment, to allow work up to +32°C room temperature
- Internal glass door, with hermetic closing (models up to -10°/-20°/-30°C do not include internal glass door)
- Solid external door with opening aids and big sized handle (double glass door optionally)
- Hermetic compressor built over dampers to reduce noise levels
- AISI 304 stainless steel internal finish
- Epoxy coated steel external finish
- Rounded corners for easy cleaning
- Heated door frame, to ensure an ice free door closing for models up to -10°/-20°/-30°C
- Access port to introduce cables and external instruments
- USB port to store system data.
- High density of 60 mm polyurethane insulation (CFC and HCFC free) high density polyurethane
- R404 (CFC and HCFC free) biodegradable cooling gas
- Magnetic gasket on the external door, to ensure a better door closing
- 4 wheels for easy movement with height-adjustable feet
- Height-adjustable perforated stainless steel shelves
- Including series 4000 controller, with graphic temperature representation
- Microprocessor (4.3'' touchscreen) for control parameters with PID system (the temperature through a NTC probe)
- Control system with battery backup and automatic recharge, for up to 48h
- The controller stores the max/min temperature and/or humidity values with their graphical representation.
- Audible and visible independent alarms for maximum and minimum temperature, with NiCd battery support backup for more than 48h of battery life (72h optionally)
- PC software
- Ethernet output
OPTIONAL ACCESSORIES
- Temperature range from +4°C to +60°C (optionally from -10/-20/-30°C to +60°C without humidity control)
- Humidity generator up to +95°C RH
- Air drying system up to 15% RH at +4°C
- Additional access ports to introduce cables and external instruments
- Thermal glass door, with heated frame in models with humidity control or lower temperatures
- External stainless steel finish
- Plugs inside of the chamber
- Potential free output for remote alarm control
- UV light tubes for sterilization
- Telescopic trays to locate instruments
- Printer for temperature and humidity data
- GSM phone alarm module
- Battery backup
Data sheet
Airflow | Horizontal without humidity |
---|---|
Temperature range without humidity | +4°C to +60°C (optionally from -10/-20/-30°C to +60°C without humidity control) |
Accuracy of display without humidity | +/-0.1 °C |
Homogeneity of temperature without humidity | +/-1.0 °C |
Stability of temperature without humidity | +/-0.5 °C |
Accuracy of NTB probe without humidity | +/-0.15 °C |
Nr of shelves | 4 |
Nr of doors | 1 |
Compressor RH | 3/8 |
Usable dimensions (WxDxH) | 640x600x1350 |
External dimensions (WxDxH) | 850x800x1980 |
Item | 701.1200.03 |
Capacity | Lt 700 |