High-volume water purification systems without storage - BENCHMARK S60UV
€0.00
Tax included
BENCHMARK S60UV ensures high quality ultrapure water, in addition to reverse osmosis water. This system has a display monitoring running status, getting data downloading from SD card or through USB interface. Double wave lenght UV lamps have been chosen to do TOC analisys and to reduce bacteria.
Description
Product Details
- It is fit to produce reverse osmosis (Type III) and ultrapure (water (Type I)
- Applications: ideal for applications requiring large amounts of on-demand water; autoclave and test chamber supplying; HPLC, GC, IC, ICP, MS, AA, TOC analysis; pharmacological, molecolar biological and microbiological application
- Integrated color LCD touchscreen interface for real-time monitoring of operating status
- Plastic external structure
- 3 doors to replace cartridges and maintaining the system
- Tube and adapter meet NSF certification
- RO system with DOW’s membrane
- High pure water quality and low running cost
- Double wavelength UV lamp (185&254nm)
- Standard pretreatment
- SD card and USB interface for storing data
- 4 ultrapure cartridges
- 3 water quality sensors
- 2 flow sensors
- 2 pumps to achieve single stage RO and sanitizing system
- Level II password protects all parameter setting
- Multiple alarm function: no feed water, full water, water quality’s standard exceeding
636.0805.08
Data sheet
| Item | 636.0805.08 |
|---|---|
| External dimension (LxWxH) | 385x500x650 mm |
| Weight | 25 Kg |
| Flow rate | 2 liters/minute (with pressure tank) |
| Output | 60 liters/hour |
| Pure water outlet | 2: reverse osmosis water, ultrapure water |
| Resistivity at 25°C | 18.2 MΩ.cmxcm |
| Conductivity | 0.0549 μs/cm |
| TOC* | < 3ppb |
| Bacteria | < 0,1 cfu/ml |
| Particle (greater than 0,2 μm) | <1/ml |
| Endotoxines | N/A |
| RNases | N/A |
| DNases | N/A |
| Ion rejection rate | 96-99% new membrane RO |
| Organic rejection rate | > 99% when MW > 200 Dalton |
| Particle and bacteria rejection rate | > 99% |
| Feed water requirements | Tapwater,temperature:5-45°C,pressure:1.0-4.0Kgf/cm x cm |
| Power | 120 W |
| Standard configuration | Main body (Including 1 set of cartridge) |